Calibration:linear/nonlinear multi-point calibration Re-test : retest the sample automatically when the result is out of the linearty range or the sample
is not suffcient
3. Working Condition
Power Supply :~100-240V,50/60Hz ,<350W Temperature:10-35 oC
Humidity < 90%,no dew
4. Operation System
Windows XP , Windows 7 , Windows 8 , Windows 10
5. System Functions
Analysis Method:end-point,kinetic,two-point, double-reagent,double-wavelegth,multi-standard etc,open to varios reagents
6. Sample/Reagent Handling
Sample volume:1-50ul,0.5 ul step
Sample Dilution : automatic/Manual pre-dilution, with dilution ratio up to 1:100
Reagent Position : 40 pcs (With refrigerated function,2-8oC) Reagent Volume : 50-300 ,1 ul step
7. Optimum calibration cuvette
40 pcs sample position Reagent refrigerated function 40 pcs reagent postion
8. Reaction Cuvette
60pcs disposable plastic cuvettes Automatic blank testing for cuvette Colorimetry system in the pattern of light splitting behind cuvette Dismountable,easilly replacement
9. Sample/Reagent Probe
Automatic liquid level detect Collision protection function Probe internal/external washing Clot detection function
8. 7-Step Auto Cleaning System
Automatic 7-step washing
Hi-tech washing station to ensure accurate results
With low water consumption: 0,5 L/h
9. Intellective software
Automatic washing cuvette when startup & shut down Automatic detect reagent volume during every startup
Reagent plate temperature detecting when startup
10. Optimum calibration cuvette
Automatic Cuvette blank testing
Automatic wash selected cuvettes
11. Dynamic and Real-time display of running status
Real-time online monitoring running status of sample tray,reagent tray and cuvette